<- Back to digest

JOA - 2026-10-01 - Journal Article

Up to 10-Year Follow-Up of Thin Polyethylene Liners in Total Hip Arthroplasty.

Leipman JH, Salehi N, Restrepo C, Smith EB, Purtill JJ, Fillingham YA

retrospective cohortLOE IIIn = 4,951 (92 at 3.9 mm, 2,690 at 5.9 mm, 2,169 at 7.9 mm)Mean 4.5 years (3.9-mm group), 6.1 years (5.9-mm and 7.9-mm groups); maximum 12.4 years.

Topics

arthroplasty
PMID: 42069017DOI: 10.1016/j.arth.2026.04.089View on PubMed ->

Key Takeaway

Thin 3.9-mm HXLPE liners in THA demonstrated zero revisions for instability or polyethylene wear at mean 4.5-year follow-up, with equivalent HOOS-JR scores to thicker liners at 2, 5, and 10 years.

Summary Depth

Choose how much analysis to show on this article page.

Summary

This single-institution retrospective cohort examined survivorship and functional outcomes of 3.9-, 5.9-, and 7.9-mm HXLPE liners in 4,951 primary THAs for osteoarthritis over up to 10 years. No significant difference in overall revision or reoperation rates was found across liner thicknesses; the 3.9-mm group had zero revisions for instability or wear, while 7.9-mm liners had a higher instability-driven revision rate than 5.9-mm liners (0.1% vs. 0.04%). HOOS-JR was equivalent across all groups at 2, 5, and 10 years, with only a transient early advantage in the 3.9-mm group at 6 months.

Key Limitation

The 3.9-mm cohort is only 92 patients with mean follow-up of 4.5 years, providing insufficient statistical power and duration to detect low-frequency catastrophic events such as liner fracture or clinically significant polyethylene wear.

Original Abstract

BACKGROUND

Using thin polyethylene liners with large-diameter femoral heads in total hip arthroplasty (THA) may decrease the risk of instability. This study investigated the 1-to-10-year survivorship of thin highly cross-linked polyethylene (HXLPE) liners.

METHODS

This was a retrospective cohort study of patients who underwent primary THA for osteoarthritis at a single institution. Patients who received an HXLPE liner with a thickness of 3.9, 5.9, or 7.9 mm and a variety of femoral head and cup sizes were included. Surgical information, patient-reported outcome measures, and 1- to 10-year outcomes were collected. Of the 4,951 cases included, 92 utilized a 3.9-mm liner, 2,690 utilized a 5.9-mm liner, and 2,169 utilized a 7.9-mm liner, with respective mean lengths of follow-up of 4.5 years (range, 1.2 to 12.2), 6.1 years (range, 1.0 to 12.1), and 6.1 years (range, 1.0 to 12.4).

RESULTS

There was no significant difference in revisions or reoperations among the cohorts. Among those revised, 7.9 mm liners were more likely to have instability as the revision indication when compared to 5.9 mm liners (0.1 versus 0.04%). The 3.9 mm group had no revisions for instability or polyethylene wear. At six months postoperatively, the Hip Disability and Osteoarthritis Outcome Score Joint Replacement (HOOS-JR) was significantly greater in patients who had a 3.9 mm liner (P = 0.004). At two, five, and 10 years postoperatively, HOOS-JR was equivalent across groups (P = 0.760, P = 0.098, P = 0.685, respectively). At five years, the 3.9 mm group had a significantly lower improvement in HOOS-JR (P = 0.036), but this difference equalized between groups at 10 years

CONCLUSIONS

Thin and standard-size HXLPE liners in THA have similar functional and clinical outcomes up to ten years postoperatively. Liners of 3.9 mm may safely be used while providing the benefit of reduced revision for instability.