The "Rescue screw" as a freehand revision strategy for implant loosening: a technical note and comparative biomechanical study on human cadavers.

Farshad M, Alakmeh A, Rohner R, Suter M, Wanivenhaus F, Widmer J, Laux CJ

cadavericLOE Vn = 5 cadaveric lumbar spines; exact screw count per trajectory not specifiedN/A

Topics

spine
PMID: 42217585DOI: 10.1016/j.spinee.2026.05.005View on PubMed ->

Key Takeaway

In a cadaveric loosening model, rescue screws placed in a craniocaudal trajectory restored 41% greater torque than controls at cemented index levels, outperforming CBT screws, which failed to restore baseline mechanical strength (-2.7%).

Summary Depth

Choose how much analysis to show on this article page.

Summary

This study compared three revision screw strategies—larger diameter (LD), cortical bone trajectory (CBT), and craniocaudal rescue screws (RS)—in a simulated loosening model using 5 human cadaveric lumbar spines with standardized 1 mm overtapping. Overall, revision screws produced 25.9% higher maximum bending torque than contralateral controls (34.0 vs. 27.0 Nm; p=0.006). In cemented index screw scenarios, RS achieved 48.9 Nm (+41%) and LD achieved 45.1 Nm (+59.2%), while CBT screws failed to restore index-level strength at 29.4 Nm (-2.7%).

Key Limitation

The sample of only 5 cadaveric spines without reported bone mineral density quantification severely limits statistical power and prevents conclusions about performance across clinically relevant osteoporosis severity strata.

Original Abstract

BACKGROUND CONTEXT

Depending on bone mineral density, implant loosening occurs in up to 54% of cases after spinal instrumentation. This leads to compromised healing, instability and mechanical pain, often requiring surgical revision. Revision strategies strongly depend on the remaining bone stock within the pedicle and include the use of larger diameter or dual pitch screws, implant cementation, alternative screw trajectories such as the cortical bone trajectory (CBT), impaction grafting, and extension to the intact adjacent segment.

PURPOSE

This study biomechanically compares screws with a 2 mm larger diameter (LD), CBT screws, and craniocaudal "rescue screws" (RS) in the setting of loosened cemented and native index pedicle screws.

STUDY DESIGN

Biomechanical study on human cadavers.

METHODS

The vertebral bodies of 5 human cadaveric lumbar spines were dissected. After randomization and simulated screw loosening with a 1 mm larger tapping, each test trajectory was tested in alternating order on all lumbar levels. RS and LD screws were analyzed for cemented and uncemented index pedicle screws. The respective contralateral pedicle was instrumented with a conventional pedicle screw as an intrasample reference. Maximum bending load-to-failure [Nm] in flexion was recorded for all screws in a biomechanic testing machine.

RESULTS

In total, revision screws showed a significantly higher (+25.9%; p=.006) maximum torque (34.0 Nm) than controls (27 Nm). In native index screws, LD screws showed the highest median torque (35.3 Nm; +19.5%). Rescue screws showed an 11.6% torque increase (32.1 Nm) compared to contralateral controls. In cemented index screws, this effect was even greater for both LD (45.1 Nm, +59.2%) and rescue screws (48.9 Nm; +41%). CBT screws did not restore the mechanical strength of the index screw in the context of prior screw loosening (29.4 Nm; -2.7%).

CONCLUSIONS

Followed by rescue pedicle screws, LD screws showed the best mechanical performance in this cadaveric screw loosening model for cemented and native index pedicle screws. CBT screws are not useful in the context of pedicle screw loosening.

CLINICAL SIGNIFICANCE

In light of the increasing number of revision surgeries, the rescue pedicle screw offers a biomechanically robust and technically simple revision strategy using standard implants and infrastructure if larger diameter screws are no longer applicable due to significant bone loss.