JOT - 2026-10-07 - Journal Article
Measuring Knee Alignment Following Distal Femur Fracture Surgery: A Novel Deep Learning Approach.
Pereira DE, Chakladar S, Harrison NJ, Rust BD, Obey MR, McAndrew CM, Berkes MB, Wilson JL
Topics
Key Takeaway
A deep learning system measured coronal knee alignment after distal femur fracture fixation with mean absolute errors of 1.1° (JLCA), 1.5° (aFTA), 1.0° (aLDFA), and 2.3° (aMPTA) compared to surgeon-validated manual measurements.
Summary Depth
Choose how much analysis to show on this article page.
Summary
This study asked whether a machine learning model could accurately automate coronal alignment measurements from routine postoperative AP radiographs in patients ≥50 years old with OTA/AO 33A and 33C distal femur fractures. The model was trained and tested on 287 radiographs, measuring JLCA, aFTA, aLDFA, and aMPTA against surgeon-validated manual annotations. Bland-Altman analysis showed 95% of differences fell within ±5° with minimal systematic bias, and Pearson correlations ranged from 0.4 to 1.0 across parameters.
Key Limitation
Pearson correlations as low as 0.4 for some parameters indicate the model performs inconsistently across alignment measures, and the clinical threshold for acceptable error in fracture patients approaching arthroplasty conversion has not been established.
Original Abstract
OBJECTIVES
To evaluate whether a machine learning-based system can accurately measure coronal knee alignment parameters from routine postoperative radiographs following surgical fixation of distal femur fractures.
METHODS
Design: Retrospective cohort study.
SETTING
Single-center academic orthopedic trauma practice.
PATIENT SELECTION CRITERIA
Adult patients 50 years of age or older who underwent surgical fixation of low-energy OTA/AO 33A and 33C distal femur fractures (including periprosthetic fractures) between November 2020 and December 2024 were retrospectively identified.
OUTCOME MEASURES AND COMPARISONS
A machine learning-based automated alignment assessment system was trained to identify standard radiographic landmarks, including the femoral intercondylar notch, femoral condyles, tibial spines, and tibial plateaus, and automatically calculate commonly used coronal alignment parameters including joint line convergence angle (JLCA), anatomic femorotibial angle (aFTA), anatomic lateral distal femoral angle (aLDFA), and anatomic medial proximal tibial angle (aMPTA). Automated measurements were compared with manual measurements annotated by a trained observer and validated by an orthopedic surgeon.
RESULTS
A total of 287 postoperative AP knee and distal femur radiographs were obtained as part of routine postoperative care and randomly divided into training and testing datasets for model development and validation. The cohort was 78.0% female with a mean age of 72.3 years (range: 50 to 97 years). Automated measurements demonstrated strong agreement with surgeon-validated measurements. Mean absolute error was 1.1° for JLCA (range: 0.0° to 3.5°), 1.5° for aFTA (range: 0.0° to 4.4°), 1.0° for aLDFA (range: 0.0° to 3.1°), and 2.3° for aMPTA (range: 0.1 to 6.4°). Pearson correlation coefficients ranged from 0.4 to 1.0, and Bland-Altman analysis demonstrated minimal systematic bias, with 95% of differences falling within a clinically acceptable range of ±5°.
CONCLUSIONS
Automated alignment assessment utilizing a machine learning-based system using routine postoperative radiographs provided accurate measurements in distal femur fracture patients and may serve as a practical adjunct for longitudinal surveillance in orthopedic trauma practice.
LEVEL OF EVIDENCE
Level III.