<- Back to digest

JOA - 2026-11-05 - Journal Article

Femoral Neck Morphology: A Computed-Tomography-Based Anatomic Framework Informs Medial Collar Design for Contemporary Collared Cementless Femoral Stems.

Kelly JJ, Helgeson EE, Alshaibi RR, Rhodes NG, Pagnano MW, Hevesi M

retrospective cohortLOE IVn = 100N/A

Topics

arthroplasty
PMID: 42988283DOI: 10.1016/j.arth.2026.10.050View on PubMed ->

Key Takeaway

At the femoral neck osteotomy plane, medial cortical thickness averaged 5.3 mm, which represents the optimal collar length to minimize overhang/underhang for collared cementless femoral stems.

Summary Depth

Choose how much analysis to show on this article page.

Summary

This study quantified femoral neck cortical thickness at and adjacent to the osteotomy plane in 100 consecutive primary THAs using preoperative CT multiplanar reconstruction to inform collared cementless stem design. Cortical thickness was measured at five regional locations at the osteotomy plane and at ±5.30 mm, revealing that medial and postero-medial cortices were thickest and increased distally (P<0.001), while antero-medial thickness plateaued at the osteotomy level (P=0.641 for distal increment). The median medial cortical thickness at the osteotomy plane was 5.3 mm, representing the collar length that minimizes mean absolute overhang or underhang.

Key Limitation

Single-surgeon, single-stem-design cohort with no demographic subgroup analysis (sex, BMI, femoral morphology by Dorr classification) prevents determination of whether the 5.3 mm collar optimum applies across the full spectrum of femoral anatomy encountered in practice.

Original Abstract

BACKGROUND

The level of femoral neck resection in total hip arthroplasty (THA) affects implant position, restoration of limb length, and degree of cortical support available for collared cementless stems. With increasing use of computed tomography (CT)-based planning and collared stems, understanding femoral neck and calcar morphology at the osteotomy plane is valuable. The aims of this study were to (1) quantify femoral neck cortical thickness at the achieved osteotomy plane reconstructed from postoperative imaging and adjacent levels, (2) assess the effect of osteotomy height on regional cortical support, and (3) determine the medial collar length that best matches cortical thickness at the osteotomy plane.

METHODS

We reviewed preoperative CT scans from 100 consecutive primary THAs performed by a single surgeon using a single cementless femoral stem design. Using multiplanar reconstruction, calcar cortical thickness was measured on axial images at the osteotomy plane (zero mm) and at standardized levels 5.30 mm proximal and 5.30 mm distal to that plane. At each level, thickness was recorded at five locations (anterior, posterior, medial, antero-medial, and postero-medial). The medial cortical thickness at the osteotomy plane was used to determine the collar length, minimizing absolute overhang/underhang, which corresponds to its median.

RESULTS

Cortical thickness increased from proximal to distal levels, most prominently in the medial (P < 0.001) and postero-medial (P < 0.001) regions. Antero-medial thickness increased from the proximal level to the osteotomy plane (P = 0.004), but did not increase further distally (P = 0.641). For this cohort and osteotomy technique, the collar length minimizing mean absolute overhang or underhang was 5.3 mm, the median medial cortical thickness at the osteotomy plane.

CONCLUSION

At and around the planned femoral neck osteotomy plane, medial and postero-medial cortices were thickest and increased in thickness distally, whereas antero-medial thickness plateaued. These data provide an anatomic reference that may inform cementless femoral stem collar geometry and future implant design.