Effect of Joint Loadings, Ligament Tensions, and Total Knee Designs on Kinematics During Flexion.
Parody N, Rozell JC, Walker PS
Topics
Key Takeaway
Among four non-cruciate TKA designs tested, only the replica intercondylar (REP) design produced consistent medial-pivot kinematics under both low-load surgical and high-load functional conditions, with statistically significant differences in rotation and lateral displacement versus the medially congruent design (P=0.1 threshold).
Summary Depth
Choose how much analysis to show on this article page.
Summary
This study used a custom crouching machine with simulated musculature and collateral ligaments to compare kinematics of four non-cruciate TKA designs (UC, MC, MP, REP) across low-load and high-load conditions at three collateral ligament tension levels from 12–130° flexion. Under high functional loads, UC, MC, and MP designs produced near-parallel AP motion with minimal axial rotation, while the REP design maintained stable medial condyle position with progressive lateral condyle displacement at both load levels. Ligament tension variation (0.5x–1.5x normal) had minimal effect on kinematics across all four designs, indicating bearing surface geometry dominated motion control in the REP.
Key Limitation
The simulator lacks a patellofemoral articulation, quadriceps wrap angle variation, and posterior capsule, all of which materially influence in vivo femoral rollback and axial rotation patterns.
Original Abstract
BACKGROUND
Reproducing anatomic kinematics during flexion in a total knee arthroplasty (TKA) requires a stable medial side and progressive posterior displacement of the lateral femoral condyle. Under surgical conditions, however, the forces across the knee are relatively small, and it is not easy to predict functional kinematics where loads are much higher. The aims of this study were to determine the relative effects of the ligament tensions and the TKA itself on the kinematics that occur under low and high load conditions during a full range of flexion.
METHODS
A crouching machine was developed with simulated femur, tibia, muscles, and collateral ligaments. There were four non-cruciate total knee arthroplasties that were designed: ultracongruent (UC), medially congruent (MC), medial pivot (MP), and replica intercondylar (REP), with guiding surfaces. There were two loading conditions: low load to simulate surgical conditions and high load for functional conditions. There were three collateral ligament tensions: normal, 0.5x normal, and 1.5x normal. The motion of the femur on the tibia was measured from 12 to 130° flexion. The transverse femoral axis projected on the tibia demonstrated displacements of the lateral and medial femoral condyles, as well as axial rotation. Representative pilot statistical tests were carried out for the MC and REP designs, with P = 0.1.
RESULTS
Under low load, all designs showed small medial displacements, but with some lateral displacements and axial rotations. Under high load, the UC, MC, and MP showed almost parallel motion with small anterior-posterior (AP) displacements and rotations. The REP showed stable medial and more pronounced lateral displacements for both low and high loads. For all four designs, the behaviors were similar for the three ligament tension conditions. There was a statistically significant difference between rotations and lateral displacements of the MC and REP under low and high load.
CONCLUSIONS
The kinematics under low-load surgical conditions generally did not match the motion under functional conditions, except for the REP design, which produced similar motion for surgical low loads and functional high loads. For the REP design, the motion was controlled mainly by the contoured shapes of the bearing surfaces.