JOA - 2026-11-14 - Journal Article
Outcomes of Rotating Hinge Knee Implants in Revision Total Knee Arthroplasty: Instability as a Potential Risk Factor for Mechanical Failure.
Vincent A, Servien E, Lustig S, Batailler C
Topics
Key Takeaway
Rotating hinge knee implants in revision TKA achieved 79.2% all-cause survival at 5 years and 67.3% at 10 years, with instability as the index indication independently predicting mechanical failure (p=0.019).
Summary Depth
Choose how much analysis to show on this article page.
Summary
This single-center retrospective study evaluated complications, survival, and failure risk factors in 230 patients undergoing revision TKA with rotating hinge knee implants over an 11-year period. Kaplan-Meier analysis demonstrated all-cause survival of 79.2% at 5 years and 67.3% at 10 years; mechanical revision survival was 85.0% at 5 years and 74.7% at 10 years. Multinomial logistic regression identified instability as the index revision indication (p=0.019) and BMI ≥30 (p=0.014) as independent risk factors for reoperation.
Key Limitation
Single-center retrospective design introduces selection bias in RHK implant indication, and the heterogeneous indications (instability, bone loss, infection, fracture) limit the ability to isolate instability-specific failure mechanisms from confounding surgical complexity.
Original Abstract
PURPOSE
Revision total knee arthroplasties (RTKA) with rotating hinged knee (RHK) implants are increasingly used in complex cases. Despite advances in implant design, the complication rate remains a major concern regarding RHK implants. This study aimed to assess the complications, revisions, survival rates, and potential risk factors of failure after RTKA with RHK at a minimum follow-up of two years.
METHODS
This retrospective, single-center study included 230 patients who underwent RTKA with RHK implants between 2010 and 2021. An RHK was implanted in cases of collateral ligament insufficiency, severe bone loss, some periprosthetic fractures, severe stiffness, or extensive debridement for infection. The mean age was 69 years (range, 43 to 92). The mean body mass index (BMI) was 30.8 (range, 13.8 to 54.7). Survival rates were analyzed using Kaplan-Meier methods based on the following endpoints: all-cause and mechanical cause revisions. Multinomial logistic regression was used to identify risk factors for implant failure. The mean follow-up was 55.2 months (range, 24.6 to 166.8).
RESULTS
Aseptic loosening was observed in 6.1% of cases (n = 14), dislocation in 3% (n = 7), and extensor mechanism rupture in 3.0% (n = 7). The overall implant survival rates were 79.2% at five years and 67.3% at 10 years for all-cause revisions and 85.0% at five years and 74.7% at 10 years for mechanical revisions. The instability as a cause of RTKA was a significant risk factor for mechanical revision (P = 0.019). A BMI below 30 was associated with a decreased risk of reoperations (P = 0.014). There were no significant risk factors that were identified for all-cause revisions.
CONCLUSION
The RTKA with RHK implants provided acceptable survival rates, but with a high risk of complications. An RTKA performed for instability represents a particularly high-risk scenario, requiring optimized surgical planning and implant selection to minimize the risk of mechanical failure.