<- Back to digest

JOA - 2026-11-12 - Journal Article

Computer-Assisted Navigation Combined with Three-Dimensional-Printing Custom Augments for Personalized Bone Defect Restoration and Acetabular Reconstruction in Complex and Revision Total Hip Arthroplasty.

Zheng Y, Geng X, Zhang Y, Li Z, Li F, Zhao M, Li Y, Wang X, Dong Z, Tian H

prospective cohortLOE IVn = 34Not reported beyond perioperative period

Topics

arthroplasty
PMID: 43020153DOI: 10.1016/j.arth.2026.11.045View on PubMed ->

Key Takeaway

A CT-based navigation system with 3D-printed custom augments achieved moderate-to-strong correlation between planned and executed acetabular angles (anteversion r=0.772, abduction r=0.535) and corrected leg length discrepancy from median -12mm to 0.5mm in 34 complex/revision THA cases.

Summary Depth

Choose how much analysis to show on this article page.

Summary

This prospective case series evaluated a novel Visual Treatment Solution (VTS) CT-based navigation system combined with 3D-printed custom augments for acetabular reconstruction in 34 complex primary and revision THA cases performed August 2023–August 2025. Primary outcomes were accuracy of postoperative acetabular anteversion and abduction versus preoperative plan; secondary outcomes included offset restoration, LLD correction, and functional scores. Anteversion correlation was r=0.772 and abduction r=0.535 versus plan; LLD corrected from median -12mm to 0.5mm; HHS, WOMAC, and SF-12 all improved significantly (P<0.001), with one intraoperative femoral fracture.

Key Limitation

The absence of a control group and the very short, undefined follow-up period make it impossible to determine whether improved component positioning translates into reduced revision rates or superior long-term implant survival compared to conventional techniques.

Original Abstract

BACKGROUND

The management of bone defects and the restoration of acetabular stability remain technically demanding in complex primary and revision total hip arthroplasty (THA). Computed tomography (CT)-derived preoperative planning in conjunction with three-dimensional (3D)- printing technology enables personalized bone defect restoration and acetabular reconstruction. We established a novel Visual Treatment Solution (VTS) navigation system that allows real-time tracking and guidance of acetabular reaming as well as the insertion of custom augments and the acetabular cup, thereby translating the individualized virtual plan into precise surgical execution. The purpose of this study was to evaluate the accuracy and efficacy of the navigation system in complex primary and revision THA.

METHODS

This prospective case series enrolled 34 patients who underwent complex primary or revision THA between August 2023 and August 2025. All surgeries were performed with the assistance of the navigation system. Primary outcomes included the accuracy of acetabular anteversion and abduction angles compared to preoperative plans. Secondary outcomes comprised femoral offset, acetabular offset, vertical center of rotation, leg length difference (LLD), functional assessments (Harris Hip Score [HHS], Western Ontario and McMaster Universities Osteoarthritis Index [WOMAC], 12-Item Short-Form Health Survey [SF-12]), and perioperative complications.

RESULTS

Strong to moderate correlations were observed between preoperative planned and postoperative acetabular angles (anteversion: r = 0.772, P < 0.001; abduction: r = 0.535, P = 0.001). Leg length difference was corrected postoperatively, from a median of -12 to 0.5 mm (P = 0.001). All functional scores improved substantially following surgery (P < 0.001 for HHS, WOMAC, and SF-12). Perioperative complications were minimal: only one case of femoral fracture occurred due to underlying osteoporosis.

CONCLUSIONS

The navigation system is a suitable option for complex primary or revision THA, particularly in cases involving bone defects in the acetabular side. Personalized preoperative planning is essential to achieve good radiograph and clinical outcomes in these cases.