Newer Generation DePuy INHANCE™ Modular Anatomic Convertible Polyethylene Glenoid and Platform Baseplate have Improved Resistance to Micromotion and Dissociation at the Polyethylene-Metal Junction.
Kane LT, Myers C, Smith M, Cuff D, Jawa A, Wiater JM, Austin L
Topics
Key Takeaway
The DePuy INHANCE™ modular glenoid demonstrated 55% greater resistance to shear dissociation (1104N vs. 713N) and 105% greater resistance to axial dissociation (512N vs. 250N) compared to a competitive modular platform system.
Summary Depth
Choose how much analysis to show on this article page.
Summary
This study biomechanically evaluated the DePuy INHANCE™ modular anatomic convertible glenoid PE liner and platform baseplate under worst-case loading conditions per ASTM F028-17 standardized testing. The INHANCE™ system produced significantly less tensile displacement than an all-polyethylene glenoid (0.1 mm vs. 0.4 mm; p=0.03) and significantly higher dissociation thresholds under both shear (1104N vs. 713N; p<0.001) and axial loading (512N vs. 250N; p<0.001) compared to a competitive modular system. These results suggest the INHANCE™ PE-metal locking mechanism represents a mechanical improvement over prior metal-backed and competing modular glenoid designs.
Key Limitation
No in vivo or cadaveric validation is provided, so whether the ASTM F028-17 worst-case loading conditions accurately replicate the physiologic forces and failure modes seen clinically remains unestablished.
Original Abstract
INTRODUCTION
Metal-backed glenoid (MBG) designs in anatomic total shoulder arthroplasty (aTSA) were introduced in the 1980s to address concerns with the high rate of loosening of all-polyethylene glenoid components, offering potential advantages of durable glenoid fixation and easy revisability, but have historically demonstrated concerning rates of mechanical failure including polyethylene (PE) liner dissociation. Recently, platform baseplates have been introduced which can be used in anatomic or reverse TSA configurations. Platform baseplates may improve clinical performance, but more biomechanical analysis is needed to support their implant design improvements over the failed MBG designs of the past. The purpose of this study was to biomechanically evaluate a newer generation DePuy INHANCE™ modular anatomic convertible glenoid PE liner and platform baseplate and its ability to withstand mechanical forces and resist polyethylene dissociation when subjected to worst-case loading conditions.
METHODS
The DePuy INHANCE™ convertible glenoid + platform baseplate were tested for resistance to micromotion from repetitive loading and compared to an all-polyethylene glenoid component. The study implant was also evaluated for its ability to withstand tensile displacement and polyethylene dissociation from shear and axial forces and compared against another commonly used modular anatomic glenoid PE liner and platform baseplate implant. Micromotion testing was performed in accordance with the standardized modular dissociation test method (ASTM F028-17) that was developed specifically for modular anatomic glenoid systems.
RESULTS
The average tensile displacement of the INHANCE™ study implant after cyclic rocking was significantly less than the all-polyethylene glenoid component (0.1 mm vs. 0.4 mm; p=0.03). The INHANCE™ polyethylene-metal junction had a significantly greater resistance to modular dissociation under shear loading (1104N, 95% CI [1068, 1140] vs. 713N, 95% CI [687, 739]; p<0.001) and axial loading (512N, 95% CI [493, 531] vs. 250, 95% CI [205, 295]; p<0.001) than was demonstrated by a competitive modular anatomic glenoid PE liner and platform baseplate system.
CONCLUSION
Our investigation shows that the DePuy INHANCE™ modular anatomic convertible glenoid PE liner and platform baseplate demonstrates favorable biomechanical performance against micromotion and polyethylene dissociation. These results should be considered when evaluating the current and future role of platform baseplates in aTSA.